The FT/IR-6000 Series FTIR spectrometers offer an optical bench which can be customized for use with the widest range of experimental requirements, including full vacuum, step-scan, microscope imaging and more.
The FT/IR-6000 optical bench is based on a 28 degree retro-reflector (corner-cube) mirror Michelson interferometer and designed for absolute flexibility; in its ‘standard’ configuration it has an extensive range of source, beam splitter and detector options from Near-IR to Mid-IR to Far-IR measurements
The FT/IR-6000 is purgeable as standard with an option for full or partial vacuum, to completely eliminate the interference of environmental gases and vapors on the sample spectrum (CO2 and H20 elimination algorithms are also provided). As an option, the interferometer, sample chamber and detector areas can all be independently evacuated (as can the IRT-5000/7000 FTIR microscopes)to completely eliminate gases and vapors for the most sensitive sample measurements. The FT/IR-6000 optical benches can also be used in a combination of vacuum and purge for faster sample turn-around.
The FT/IR-6000 has several options for emission port optics with a range of window materials for measurement of external light sources. These can be coupled with customizable collection optics to focus a range of sizes and power outputs of light sources into the interferometer.
The FT/IR-6000 comprises three basic optical benches utilizing the same Spectra Manager II software platform. Models range from the FT/IR-6600 with a resolution of 0.4 cm-1(S/N 45,000:1), and the FT/IR-6700 with a resolution of 0.25cm-1 (S/N 47,000:1) both with aluminum optics, to the FT/IR-6800 with 0.07cm-1 resolution (S/N 55,000:1)utilizing gold optics.
All instruments includea high output ceramic source, KBrbeamsplitter and DLATGS detector.
The real power of these systems is not only the uncompromising design and the high optical throughput, but also the customization options for virtually any experimental requirement
The FT/IR-6000 Series is controlled by Spectra Manager™ Suite cross-platform software. This includes a wide variety of standard functions such as spectra measurement, Quick-Start, QC Compare and quantitative analysis. Additional applications can be added, such as chemometrics (PLS, PCR, CLS etc.). The sample measurement screen can be fully customized according to user requirements and each customized screen and parameters can be saved for future use (User Adaptive Software capability).
Model | FT/IR-6600 | FT/IR-6700 | FT/IR-6800 |
---|---|---|---|
Standard Wavenumber Measurement Range | 7,800 to 350 cm-1 | ||
Optional Extended Wavenumber Range | 25,000 to 10 cm-1 | ||
Display Wavenumber Range | 15,000 to 0 cm-1 (standard) 25,000 to 0 cm-1 (optional) |
||
Wavenumber Accuracy | Within ± 0.01 cm-1 (theoretical value) | ||
Maximum Resolution | 0.4 cm-1 0.07 cm-1 (optional) |
0.25 cm-1 0.07 cm-1 (optional) |
0.07 cm-1 |
Optical System | Single beam | ||
Sample Chamber | Size: 200 mm (W) × 260 mm (D) × 185 mm Optical path: Center focus, light axis 70 mm high |
||
Interferometer | 28º Michelson interferometer Corner cube mirror interferometer, with auto-alignment mechanism, sealed structure, DSP control |
||
N2 Purge | Interferometer, Sample Compartment, Detector | ||
Vacuum Instrument | Full and Partial options | ||
Mirror Coating | Aluminium | Aluminium | Gold |
Drive Method | Mechanical bearing, electromagnetic drive |
||
Drive Speed | 0.5, 1, 2, 3, 4, 5, 6, 7, 8 mm/sec AUTO DLaTGS 2.0 mm/sec. MCT (optional) 4.0 mm/sec |
0.5, 1, 2, 3, 4, 5, 6, 7, 8 mm/sec AUTO DLATGS 2.0 mm/sec. MCT (optional) 4.0 mm/sec |
0.125, 0.25, 0.5, 1, 2, 3, 4, 5, 6, 7, 8 mm/sec AUTO DLATGS 2.0 mm/sec. MCT (optional) 4.0 mm/sec |
Rapid Scan | 20 Hz (optional) | 20 Hz (standard) | |
Beam Splitter | Standard: Ge/KBr Option: Quartz, Si/CaF2, Ge/CsI, Mylar (interchangeable) |
||
Replacement Method | Secure-lock beamsplitter catch system (Option: Automatic beam splitter exchange system) |
||
Light Source | Standard: High-intensity ceramic source Option: Halogen lamp, water-cooled mercury light source Up to three light sources may be installed simultaneously including external light sources |
||
Detector | DLaTGS (with Peltier temperature control) (standard) | ||
Optional Detectors | W-MCT, M-MCT, N-MCT, Si, InSb, InGaAs, PAS, Si bolometer (optional) Two detectors may be mounted simultaneously within the instrument. Up to two external detectors may be installed. |
||
Signal-to-Noise Ratio: (4 cm-1, 1 min, near 2,200 cm-1) |
45,000:1 | 47,000:1 | 55,000:1 |
Gain Switching | AUTO, 1, 2, 4, 8, 16, 32, 64, 128 | ||
100%T Line Flatness | Within 100 ± 1.0%T (4,000 to 700 cm-1, continuous repetitive measurement) |
||
Communication | USB 2.0 | ||
FTIR Main Instrument | Dimensions: 600 (W) × 670 (D) × 315 (H) mm, Weight:56 kg |
||
Power Supply | Dimensions:200 (W) × 285 (D) × 90 (H) mm, Weight:4.7 kg Unit can be placed on its base or on its side |